Biological Safety Cabinet, Class II Type B2

bioMARK-II-B2 – Class II Type B2 Biohazard Safety Cabinet (Total Exhaust Cabinet)

The bioMARK-II-B2 is a Class II Type B2 Biohazard Safety Cabinet, also known as a Total Exhaust Cabinet, designed to provide maximum personnel, product, and environmental protection. Unlike other biosafety cabinets, the B2 model features a 100% exhaust system with no air recirculation, ensuring the highest level of containment for hazardous biological agents and chemical vapors. It creates a Class 100 ultra-clean work zone with vertical laminar airflow, making it ideal for applications requiring absolute contamination control.

Key Features & Benefits

  • ✔ NSF Standard 49 Compliant Filtration System – Ensures superior air purification and compliance with global safety standards.
  • ✔ Total Exhaust System (100% HEPA Filtered) – No recirculation, eliminating contamination risks.
  • ✔ Maximum Containment – All contaminated plenums are surrounded by negative pressure, virtually eliminating leakage.
  • ✔ Balanced, Low-Noise Blowers – Engineered for continuous operation with optimized motor capacity.
  • ✔ Durable & Hygienic SS304 Construction – Seamless work zone with minimal welded joints prevents contaminant buildup and rusting.
  • ✔ Adjustable Front Sash Opening (8") – Includes alarm system to prevent over-opening and maintain containment.
  • ✔ UV Germicidal Lamp (15W or 30W) – Provides effective preliminary sterilization.
  • ✔ Ergonomic Illumination – Electronically ballasted fluorescent lighting enhances operator comfort and visibility.
  • ✔ Aesthetic & Robust Design – Powder-coated GI body in dual-color finish (fully SS304 construction available upon request).
  • ✔ Exhaust HEPA Module – Positioned above the false ceiling or outdoors, allowing easy access for maintenance.
  • ✔ Comprehensive Testing & Calibration – Factory and on-site testing for HEPA filter integrity, cleanliness, and airflow validation. All instruments are calibrated with NABL or International Standards traceability.

Optional Features

  • ✔ Virus Burnout Unit – Additional sterilization for maximum safety.
  • ✔ Second Stage Exhaust HEPA Filtration – Provides enhanced air purification to meet advanced safety requirements.

Critical Considerations for Installation

  • ✔ Compensating Fresh Air Supply – Since the biosafety cabinet exhausts conditioned air, a Positive Pressure / Fresh Air Module is recommended to maintain a stable lab environment.
  • ✔ Air Conditioning Requirements – For every 200 CFM of fresh air added, 1 ton of AC should be installed. A 3' Type B2 cabinet requires a minimum of 2 TR AC.
  • ✔ Heat Compensation for Exhaust Air – Exhaust air at 450 CFM would require 15 kW of heating to maintain 80°C, which is energy-intensive. To avoid this, an additional HEPA filtration system with a blower is recommended, following USA & European standards.
  • ✔ HEPA Filter Integrity Testing – All HEPA filters must be tested using an Aerosol Photometer post-installation to ensure environmental and operational safety.

Warranty & Additional Services

  • ✔ 12-Month Warranty (excluding consumable parts and accessories)
  • ✔ IQ/OQ Documentation Available Upon Request

Additional Information

  • ✔ Production Capacity: 200 units per annum
  • ✔ Delivery Time: 6 week

Product Details:

Minimum Order Quantity 1 Unit
Cabinet Material Mild Steel, Stainless Steel
Main Filter HEPA Filter
Sterilization UV Germicidal Tube
Electric Supply less than 1kW, Single Phase
Noise Level Less than 65 db
Working area size 3 ft, 4ft, 6ft Wide X 2ft deep X 2'3" high

 